Processing Record for this file: (Make notes here unique for this file) Loading 7/1 - 8/1/2014 Data loaded from detarred, dezipped files from the DMF To reduce Operator need for frequent filter changes, dilution turned on @ 14-02-21 21:12. Every 24-h, the dilution is turned off for 5 minutes. This was not long enough to get an undiluted sample. As always, data not reported for T(blue) < 0.7 Impactor in place. As confirmation of correct signal polarity, the flow signal decreases slightly when the impactor is in the 1-um state. Some evidence of supermicron particles. A number of local events resulting in high absorbance. These were not edited. 2014-07-02 03:10, 2014-07-02 05:45 DQR D141203.8 Appears that Operator changed filter @ 03:15 and did not reset transmission until 04:15. Then there was another transmittance discontinuity at 05:20. All data should be considered incorrect. 2014-07-19 00:05:11, 2014-07-20 15:26:16 DQR D141203.9 Pentras flow oscillating. Data should be considered incorrect. 2014-07-22 14:00:24, 2014-07-24 22:22:56 DQR D141203.10 Impactor appears to not be operating. Data Gaps > 1 h DQR D141203.11 2014-07-01 07:21:01, 2014-07-01 14:26:30 2014-07-02 03:04:10, 2014-07-02 05:31:51 2014-07-03 01:54:40, 2014-07-03 04:56:29 2014-07-03 10:47:56, 2014-07-03 13:25:29 2014-07-05 09:32:06, 2014-07-05 13:54:28 2014-07-07 22:58:44, 2014-07-08 05:38:27 2014-07-08 15:17:25, 2014-07-08 19:33:27 2014-07-09 07:24:53, 2014-07-09 12:57:27 2014-07-11 07:11:43, 2014-07-11 15:42:30 2014-07-15 06:58:35, 2014-07-15 14:21:25 2014-07-15 18:13:58, 2014-07-15 19:26:25 2014-07-15 21:07:58, 2014-07-16 05:29:25 2014-07-17 11:14:50, 2014-07-17 16:15:24 2014-07-18 15:33:04, 2014-07-20 12:39:23 2014-07-21 03:00:01, 2014-07-21 04:58:23 2014-07-22 06:51:15, 2014-07-22 13:11:23 2014-07-25 10:20:18, 2014-07-25 12:26:22 2014-07-27 10:46:55, 2014-07-27 16:58:21