20-jul-19 08:14:54 SERIOUS INSTRUMENT PROBLEM !!! Problem with air flow pump or plumbing. Flow < 1 vLPM. --- 20-jul-19 08:14:54 Measurements ended --- . Number of lamps (L) = 7 . Filter running time = 13.66 hours . Total airflow this filter = 1.35 cubic meters . Mean BC concentration of all lamps = 430 ng/vm3 . Report for lamp 1 (370 nm): . Ending voltages: sen(beam) = 1.249 sen(zero) = 0.021 . ref(beam) = 2.346 ref(zero) = 0.021 . Optical attenuation of filter deposit = 50 . Total aerosol black carbon on filter = 633 ng Mean aerosol black carbon concentration = 471 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 161 ppm. --- Report for lamp 2 (470 nm): . Ending voltages: sen(beam) = 1.062 sen(zero) = 0.021 . ref(beam) = 2.511 ref(zero) = 0.021 . Optical attenuation of filter deposit = 38 . Total aerosol black carbon on filter = 626 ng Mean aerosol black carbon concentration = 466 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 109 ppm. --- Report for lamp 3 (520 nm): . Ending voltages: sen(beam) = 1.502 sen(zero) = 0.021 . ref(beam) = 1.770 ref(zero) = 0.021 . Optical attenuation of filter deposit = 33 . Total aerosol black carbon on filter = 587 ng Mean aerosol black carbon concentration = 437 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 120 ppm. --- Report for lamp 4 (590 nm): . Ending voltages: sen(beam) = 1.086 sen(zero) = 0.021 . ref(beam) = 2.350 ref(zero) = 0.021 . Optical attenuation of filter deposit = 28 . Total aerosol black carbon on filter = 580 ng Mean aerosol black carbon concentration = 431 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 350 ppm. --- Report for lamp 5 (660 nm): . Ending voltages: sen(beam) = .373 sen(zero) = 0.021 . ref(beam) = .790 ref(zero) = 0.021 . Optical attenuation of filter deposit = 25 . Total aerosol black carbon on filter = 578 ng Mean aerosol black carbon concentration = 430 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 176 ppm. --- Report for lamp 6 (880 nm): . Ending voltages: sen(beam) = .773 sen(zero) = 0.021 . ref(beam) = 1.158 ref(zero) = 0.021 . Optical attenuation of filter deposit = 17 . Total aerosol black carbon on filter = 536 ng Mean aerosol black carbon concentration = 398 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 235 ppm. --- Report for lamp 7 (950 nm): . Ending voltages: sen(beam) = 1.326 sen(zero) = 0.021 . ref(beam) = 1.221 ref(zero) = 0.021 . Optical attenuation of filter deposit = 15 . Total aerosol black carbon on filter = 508 ng Mean aerosol black carbon concentration = 378 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 262 ppm. --- Disk space = 2.540012E+08 bytes free: One entry at every 10 minutes. There is space for 107994 hours ( 4500 days) more data ---